ICOTEC AG
2 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (2)
Devices registered here (2)
- Spinal Vertebral Body Replacement Device - CervicalProduct code PLR (opens in a new tab) · Class II · Orthopedic · 21 CFR 888.3060
- Spinal Vertebral Body Replacement DeviceProduct code MQP (opens in a new tab) · Class II · Orthopedic · 21 CFR 888.3060
Recalls & enforcement
- Class IIOngoing2023-11-28icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was…
- Class IITerminated2022-11-03During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “ICOTEC AG (FEI 3006493760) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3006493760-icotec-ag/
Alerts for ICOTEC AG
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.