RegWardenFDA facility compliance records
Recall

Class II — Icotec Ag

Recalling firmicotec ag
Registered facilityICOTEC AG (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonDuring internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.
ProductVADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702
Quantity6 devices
DistributionUS Nationwide distribution in the state of GA.
Initiated2022-11-03
Classified2022-12-19
Terminated2026-09-10
Recall numberZ-0728-2023
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Icotec Ag (Z-0728-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/91174/

Not open yet

Alerts for icotec ag

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned