RegWardenFDA facility compliance records
Recall

Class II — Icotec Ag

Recalling firmicotec ag
Registered facilityICOTEC AG (the firm's only registered facility)
ClassificationClass II
StatusOngoing
Reasonicotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.
ProductVADER pedicle system, torque wrench, Catalog Number 42-703
DistributionUS: AZ. MI, OH, CA, CO, TX, NY, NC, DE, PA
Initiated2023-11-28
Classified2024-01-10
Recall numberZ-0701-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Icotec Ag (Z-0701-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93577/

Not open yet

Alerts for icotec ag

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.