Instrumentation Laboratory Co.
24 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (24)
Registered facilities (3)
- INSTRUMENTATION LABORATORY CO. FEI 1217183 · MA
- INSTRUMENTATION LABORATORY CO. FEI 2431530 · NY
- INSTRUMENTATION LABORATORY CO. - Devens FEI 3037000637 · MA
Recalls & enforcement
20263 records
- Class IIOngoing2026-05-21The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalitie…
- Class IIOngoing2026-05-08Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
- Class IIOngoing2026-01-20Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requir…
20253 records
- Class IIOngoing2025-12-11Potential for microbial contamination.
- Class IIIOngoing2025-11-18Recalled lots were manufactured with double the amount of preservative concentration.
- Class IIOngoing2025-02-03Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL Liquid Anti-Xa and HemosIL Heparin Calibrator. In some complaints,…
20241 record
- Class IIOngoing2024-12-18Control indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range.
20233 records
- Class IIOngoing2023-09-29Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to fa…
- Class IIOngoing2023-07-13Sample misidentification could occur under specific conditions and patient management altered based on an incorrectly assigned result
- Class IIOngoing2023-02-03Removes the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments
20222 records
- Class IIOngoing2022-04-08Variable and out of specification QC results with HemosIL ReadiPlasTin. IL will remove all lots from the market and convert customers to an alternative prothrombin time (PT) reagent, specifically HemosIL RecombiPlasTin 2G.
- Class IIOngoing2022-02-22Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL…
20212 records
- Class IIOngoing2021-10-13The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or do not pass for each vial of reagent, there is a potential ri…
- Class IIOngoing2021-08-25Labeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days
20191 record
- Class IITerminated2019-07-23Prolonged clotting times
20181 record
- Class IITerminated2018-08-01There is a potential for increased imprecision, out of range quality controls and prolonged sample results.
20171 record
- Class IITerminated2017-08-02Instrumentation Laboratory Co. received customer reports of performance issues with some vials of Lot Nos. N1166235 and N0177760, including increased imprecision, out of range quality controls and prolonged sample results.
20151 record
- Class IITerminated2015-12-18This Lot is not meeting labeled on-board instrument stability claims of 5 days.
20142 records
- Class IITerminated2014-11-21Some vials of HemosIL PT-Fibrinogen HS PLUS, Part No. 0008469810 (various lots manufactured prior to June 2014) exhibit a yellow/brownish color after reconstitution (yellow color or darker) instead of the typical white to off-white color. Vials exhibiting this yellow/brownish c…
- Class IITerminated2014-06-03Potential for sample misidentification.
20133 records
- Class IITerminated2013-11-15HemosIL Low, Abnormal, Normal Controls UNASSAYED (no-value assigned) lacks 510(k) approval
- Class IITerminated2013-07-19Potential carryover issue that can can cause shortened APTT clotting times on the instrument.
- Class IITerminated2013-04-09Instrumentation Laboratory is recalling HemosIL RecombiPlasTin 2G (RPT2G), Part No. 0020009580 (8 mL Size) due to being Out of Specification.
20121 record
- Class IITerminated2012-08-03Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “Instrumentation Laboratory Co. — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-1217183-instrumentation-laboratory-co/
Alerts for Instrumentation Laboratory Co.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.