Class II — Instrumentation Laboratory

Recalling firmInstrumentation Laboratory Co.
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonThe firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or do not pass for each vial of reagent, there is a potential risk of reporting an erroneous result.
ProductHemosIL ReadiPlasTin (20 mL), Part number 0020301400. An in vitro diagnostic thromboplastin reagent.
Quantity5828 kits (310 US)
DistributionWorldwide distribution - US Nationwide distribution in the states of AZ, CA, IA, IL, KS, ME, MI, NC, NY, OK, SC, and VA. The countries of Canada, Chile, China, Costa Rica, Italy, Japan, and South Korea. ***Updated 3/3/22*** Expansion includes states of IN, MN, and NC.
Initiated2021-10-13
Classified2021-11-24
Recall numberZ-0276-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Instrumentation Laboratory (Z-0276-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88930/

Not open yet

Alerts for Instrumentation Laboratory Co.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.