Class II — Instrumentation Laboratory
| Recalling firm | Instrumentation Laboratory Co. |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae. |
| Product | GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010 00077430010 Product Description: The GEM Premier 7000 with iQM3 is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. |
| Quantity | 32 units |
| Distribution | Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MN, MO, MT, OH, PA, VA, WI, WV and the countries of China, Germany, Italy. |
| Initiated | 2026-05-21 |
| Classified | 2026-08-06 |
| Recall number | Z-2899-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Instrumentation Laboratory (Z-2899-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/99128/
Alerts for Instrumentation Laboratory Co.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.