Recall

Class II — Accuray Incorporated

Recalling firmAccuray Incorporated
Registered facilityAccuray Incorporated (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonIn certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
ProductCyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
Quantity292 units
DistributionWorldwide - US Nationwide and and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kenya, Korea, Republic of, Liechtenstein, Lithuania, Luxembourg, Macao, Mexico, Netherlands, New Zealand, Pakistan, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Spain, Sweden, Sw…
Initiated2026-06-16
Classified2026-07-01
Recall numberZ-2644-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Accuray Incorporated (Z-2644-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/99083/

Not open yet

Alerts for Accuray Incorporated

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.