Accuray Incorporated
17 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (17)
Registered facilities (1)
- Accuray Incorporated FEI 3003873069 · WI
Recalls & enforcement
20262 records
- Class IIOngoing2026-06-16In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housi…
- Class IIOngoing2026-06-16In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housi…
20241 record
- Class IICompleted2024-12-23Snap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right roll axes when the rings are removed or damaged.
20231 record
- Class IICompleted2023-08-03It is possible the upper roller bracket in the Radixact System gantry enclosure may not provide enough clearance for service personnel to use a jib crane for moving components while servicing the system, resulting in possible injury to the service personnel.
20211 record
- Class IITerminated2021-02-26The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.
20201 record
- Class IITerminated2020-12-10"MLC tickle error" may result in the delivered dose to effectively rotate from the planned dose.
20183 records
- Class IITerminated2018-09-26A robotics supplier notified Accuray of two manufacturing variations in the fastening of the in-line wrist and the casting of the wrist for the robot. The variation may result in premature failure of the component.
- Class IITerminated2018-08-31the Ball Cube II phantom calibration film was not manufactured within necessary dimensions and may introduce up to 0.5 mm of positional inaccuracy to the CyberKnife System
- Class IITerminated2018-04-10There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.
20171 record
- Class IITerminated2017-11-17Uncontrolled couch Z-axis movement (descent)
20163 records
- Class IITerminated2016-12-06Accuray has become aware of a potential safety issue involving the possibility of electric shock during maintenance activities from the lon Pump HV Power Supply at the back of the modulator cabinet. This issue is limited to certain G3, G4, and VSI models of the CyberKnife® Robot…
- Class IITerminated2016-11-22Accuray has become aware of a potential safety issue involving unexpected treatment robot motion when removing an accessory. This issue is limited to certain CyberKnife® Systems with software version 1 0.6.
- Class IITerminated2016-01-22The gun box mounting bracket may fail to support the weight of the gun box when in vertical (inverted) position. If this failure occurs the gun box may become loose and could come into contact with a patient.
20142 records
- Class IITerminated2014-10-31Software upgrade to correct potential safety issue related to CyberKnife System that occurs when upgrading the Treatment Delivery Software for the first generation Iris Variable Aperture Collimator.
- Class IITerminated2014-01-13Potential Safety issue with Synchrony Boom Arm Mounting Assembly - one complaint of mounting assembly detaching.
20132 records
- Class IITerminated2013-03-27A5 gearboxes responsible for pitch (head up and head down) and roll (patient left and right) motions of the RoboCouch manipulator may be defective.
- Class IITerminated2013-03-19A defect on the master-side connector block of the pneumatic tool changer could introduce a tilt in the collimator housings, resulting in a potential shift to the central axis of the radiation beam. The issue could pose significant clinical impact to plans created using the small…
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “Accuray Incorporated — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-9005157-accuray-incorporated/