Class II — Beta Bionics, Inc.

Recalling firmBeta Bionics, Inc.
Registered facilityBETA BIONICS, INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
Reasoninsulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.
ProductiLet Bionic Pancreas, REF: BB1001
Quantity1,080
DistributionUS Nationwide distribution in the states of TX, PA, AL, MI, OH, FL, TN, GA.
Initiated2026-03-12
Classified2026-04-29
Recall numberZ-1996-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Beta Bionics, Inc. (Z-1996-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98475/

Not open yet

Alerts for Beta Bionics, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.