BETA BIONICS, INC.
3 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (3)
- Inspections (1)
- Warning letters (1)
Devices registered here (1)
- Syringe, PistonProduct code FMF (opens in a new tab) · Class II · General Hospital · 21 CFR 880.5860
Recalls & enforcement
- Class IIOngoing2026-03-12insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspension…
- Class IIOngoing2025-09-23The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen transition occurs, and the user remains …
- Class IIOngoing2025-08-13it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those ico…
FDA inspections (1)
- NAI2024-10-21No Action Indicated · Compliance: Devices; Postmarket Assurance: Devices
Source: U.S. FDA Inspections Dashboard (opens in a new tab) (device inspection classifications, 2016–present). Data as of 2026-10-06. The dashboard profile can take several seconds to load.
FDA warning letters (1)
- Warning Letter2026-01-29Warning Letter (opens in a new tab) · Irvine California
Source: U.S. FDA Compliance Actions Dashboard (opens in a new tab) (device warning letters, 2016–present). Data as of 2026-10-06. Each link opens an FDA search for the letter; FDA can take several weeks to post a letter, so a very recent one may not appear yet.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “BETA BIONICS, INC. (FEI 3019004087) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3019004087-beta-bionics-inc/
Alerts for BETA BIONICS, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.