Class II — Beaver-Visitec International, Inc.

Recalling firmBeaver-Visitec International, Inc.
Registered facilityBeaver-Visitec International, Inc (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonManufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
ProductBVI Leos Laser and Endoscopy System. Model Number: OME6000U.
Quantity5 units
DistributionUS Nationwide distribution in the states of FL, MI, OK, TX, WI.
Initiated2025-12-05
Classified2025-12-31
Recall numberZ-1008-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Beaver-Visitec International, Inc. (Z-1008-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98076/

Not open yet

Alerts for Beaver-Visitec International, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.