Class II — NRT X-RAY A/S

Recalling firmNRT X-RAY A/S
Registered facilityNRT X-RAY A/S (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonX-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.
ProductIntelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system
Quantity35
DistributionUS Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV, LA, NY, PA, HI, NV, WI, MN, MI.
Initiated2024-04-17
Classified2024-05-07
Recall numberZ-1784-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — NRT X-RAY A/S (Z-1784-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/94363/

Not open yet

Alerts for NRT X-RAY A/S

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.