NRT X-RAY A/S
3 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (3)
- Inspections (1)
Devices registered here (2)
- System, X-Ray, Fluoroscopic, Image-IntensifiedProduct code JAA (opens in a new tab) · Class II · Radiology · 21 CFR 892.1650
- Interventional Fluoroscopic X-Ray SystemProduct code OWB (opens in a new tab) · Class II · Radiology · 21 CFR 892.1650
Recalls & enforcement
- Class IIOngoing2024-04-17X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is…
- Class IIOngoing2023-04-27There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.
- Class IIOngoing2022-01-21Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C, Left.
FDA inspections (1)
- VAI2023-03-24Voluntary Action Indicated · Compliance: Devices; Postmarket Assurance: Devices; Radiation Control and Health Safety Act · citations posted
Source: U.S. FDA Inspections Dashboard (opens in a new tab) (device inspection classifications, 2016–present). Data as of 2026-10-06. The dashboard profile can take several seconds to load.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “NRT X-RAY A/S (FEI 1000188474) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/1000188474-nrt-x-ray-a-s/
Alerts for NRT X-RAY A/S
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.