Class II — Sysmex America, Inc.

Recalling firmSYSMEX CORPORATION
Registered facilityView this company's 14 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusCompleted
ReasonThere is potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%, #) results obtained using the lots listed above may exhibit lower than actual PLT-F and IPF values. The issue may be recognized by a large discrepancy between the impedance platelet (PLT-I) result and the PLT-F result. In most cases, the falsely decreased PLT-F results display a PLT Abn Scattergram message with an asterisk beside the result or an Action Message: Difference between PLT and PLT- F. Check the results. These are indicators of unreliable data. In rare instances, platelet flags may be absent. The PLT-I results are unaffected.
ProductSysmex Fluorocell(R) PLT Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers
Quantity23,239 units (9353 US)
DistributionWorldwide Distribution - US (nationwide including Puerto Rico) Internationally to New Zealand, Brazil, India, Philippines, Nicaragua, Argentina, Australia, Chile, Viet Nam, Colombia, Pakistan, Malaysia, Dubai, Germany, Hong Kong, Bangladesh, Uruguay Thailand, China, Saudi Arabia, Taiwan, Honduras, Ecuador, Kuwait, South Korea, and Costa Rica
Initiated2017-09-08
Classified2017-10-24
Recall numberZ-0045-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Sysmex America, Inc. (Z-0045-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/78134/

Not open yet

Alerts for SYSMEX CORPORATION

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.