SYSMEX CORPORATION
13 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (13)
Registered facilities (14)
- SYSMEX CORPORATION TECHNO PARK FEI 3002808301 · JP
- SYSMEX CORPORATION FEI 3010421384 · JP
- SYSMEX CORPORATION i-SQUARE FEI 3010960914 · JP
- Sysmex America, Inc. - Vernon Hills FEI 3020704087 · IL
- Sysmex America, Inc. FEI 3009711478 · IL
- Sysmex Europe SE FEI 3005626422 · DE
- SYSMEX CORPORATION ONO FACTORY FEI 3002808299 · JP
- SYSMEX CORPORATION KAKOGAWA FACTORY FEI 3002759288 · JP
- SYSMEX DO BRASIL IND E COM LTDA FEI 3008526732 · BR
- Sysmex RA Co., LTD, Main Office FEI 3010188766 · JP
- SYSMEX CORPORATION SEISHIN FACTORY FEI 3007701696 · JP
- SYSMEX AMERICA, INC. FEI 1000515253 · IL
- SYSMEX REAGENTS AMERICA, INC. FEI 3006949423 · IL
- Sysmex America, Inc. FEI 3009495621 · IL
Recalls & enforcement
20252 records
- Class IIOngoing2025-12-15Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.
- Class IIOngoing2025-01-06Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.
20203 records
- Class IITerminated2020-07-06Insufficient amount of antibody without an error message or alarm
- Class IITerminated2020-03-09Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove slides or slide magazines from the CF-70, the operation of the CF-70 does not halt.
- Class IITerminated2020-01-21When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.
20181 record
- Class IITerminated2018-07-27The fluorescent platelet counts (PLT-F) and immature platelet fraction (IPF) values can be falsely low when using affected lots of Fluorocell PLT.
20172 records
- Class IICompleted2017-09-08There is potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%, #) results obtained using the lots listed above may exhibit lower than actual PLT-F and IPF values. The issue may be recognized by a large discrepancy between the impedance platelet (PLT-I…
- Class IITerminated2017-03-23Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat genera…
20161 record
- Class IITerminated2016-07-01There is a potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%) results performed using the recalled lots may exhibit lower than expected PLT-F and IPF values. The issue may be recognized by a large discrepancy between the Impedance platelet (PLT-I) v…
20141 record
- Class IITerminated2014-12-18There is a potential for elevated hemoglobin (Hgb) and impedance platelet count (PLT-I) values.
20132 records
- Class IITerminated2013-05-15The data printed on the thermal printer may omit a digit or decimal point in the results or sample identification number.
- Class IITerminated2013-03-31The current Century Break patch did not contain the same setting for the Century Break year for both the Red Hat Linux and GUI files. Due to the mismatch in file settings, the software could not properly calculate a purge date for the database files based on the customer data ret…
20121 record
- Class IITerminated2012-06-01There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “SYSMEX CORPORATION — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-9014170-sysmex-corporation/
Alerts for SYSMEX CORPORATION
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.