RegWardenFDA facility compliance records
Recall

Class II — Edwards Lifesciences, LLC

Recalling firmEdwards Lifesciences LLC
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonHoles in the sterile barrier pouch may be caused by handling of the device during manufacturing or once pouches have been removed from the unit box and stored on the shelf by customers.
ProductEdwards Lifesciences Crimpers, Model 9100CR26. The Crimpers are used to crimp the transcatheter heart valve onto the delivery system prior to use.
Quantity6,117 units
DistributionNationwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI.
Initiated2013-03-26
Classified2013-04-17
Terminated2013-09-16
Recall numberZ-1135-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Edwards Lifesciences, LLC (Z-1135-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/64747/