RegWardenFDA facility compliance records
Facility

ORASURE TECHNOLOGIES, INC.

7 recall(s) on record — most severe on record: Class I.

Activity by year

FDA actions on record by year2016: 3 recalls; 2017: 1 recalls; 2020: 2 recalls; 2021: 1 recalls320161201720182019220201202120222023202420252026
  • Recalls (7)

Devices registered here (1)

Recalls & enforcement

  • Class IIITerminated2021-02-05
    Incorrect lot of Substrate Reagent (TMB) may be included in the kit. The original kit box labeling and the specification sheet listed the Substrate Reagent (TMB) lot 6676239, but contained Substrate Reagent (TMB) lot 6679623, should not affect performance.
  • Class IIITerminated2020-07-15
    Unit boxes not properly sealed
  • Class IIITerminated2020-07-13
    Incorrect lot number for Standard Oral Fluid Negative Calibrator on specification sheet included in kit and/or additionally same error on unit box
  • Class IIITerminated2017-12-20
    The controls and calibrators contained within the cocaine test kit may cause false positive results with the negative control.
  • Class IIITerminated2016-11-10
    Shipping error. Cocaine metabolite Microplate distributed instead of Barbiturate Microplate due to incorrect selection of Microplate during manufacturing.
  • Class ITerminated2016-07-20
    Failed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C conditions.
  • Class IIITerminated2016-01-15
    OraSure Technologies, Inc. discovered the package insert included with the OraQuick HCV Rapid Antibody test Visual Reference Panel may be incorrect due to a complaint received from one of their customers. The customer reported they received the Ora Quick Ebola Visual Reference Pa…

Related

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)

Cite this record

RegWarden. “ORASURE TECHNOLOGIES, INC. (FEI 3022474803) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3022474803-orasure-technologies-inc/

Not open yet

Alerts for ORASURE TECHNOLOGIES, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.