Class I — OraSure Technologies, Inc.

Recalling firmORASURE TECHNOLOGIES, INC.
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusTerminated
ReasonFailed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C conditions.
ProductOraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola viruses.
Quantity3,075 units
DistributionDistributed to one customer only in US state of Georgia.
Initiated2016-07-20
Classified2016-08-17
Terminated2017-05-10
Recall numberZ-2403-2016
InitiationVoluntary: Firm initiated