COOPER MEDICAL, SRL
57 recall(s) on record — most severe on record: Class I.
Activity by year
- Recalls (57)
Devices registered here (1)
- Media, ReproductiveProduct code MQL (opens in a new tab) · Class II · Obstetrics/Gynecology · 21 CFR 884.6180
Recalls & enforcement
20261 record
- Class IOngoing2026-08-11Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
20251 record
- Class IIOngoing2025-06-11The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.
20243 records
- Class IIOngoing2024-11-15The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
- Class IIOngoing2024-07-31During the manufacturing process the devices were laser-marked with the incorrect part number.
- Class IIOngoing2024-02-16Low concentration of sodium pyruvate in the wash could compromise gamete development
20233 records
- Class IIOngoing2023-12-20The firm has become aware of a sudden increase in complaints for three associated lots. Performance issues may lead to impaired embryo development prior to the blastocyst stage.
- Class IIOngoing2023-08-29Incorrect component descriptions within the "Contents" section on the Tyvek lid for individual kit trays.
- Class IIOngoing2023-02-13It has come to CooperSurgical's attention that the affected Product may contain a medium other than the Global Total LP single step medium which does not contain the protein source required for embryo culture and development.
20226 records
- Class IIOngoing2022-12-15The affected product may contain a medium other than the Biopsy Medium. There is a risk that use of the Product from this lot could cause degradation of the embryo.
- Class IIOngoing2022-10-20Firm has received 11 complaints involving a loose or detached metal clips, including 3 cases where the metal clip was ingested by the patient, and 3 cases where the metal clip was removed from the patient's mouth. In one case, the patient suffered a laceration to the throat.
- Class IIOngoing2022-08-12There is a potential breach to the sterile barrier containing the Humidifier Bottles due to the packaging of the filter box located inside the bottle packaging. Damage to the sterile barrier may result in use of an unsterilized device, which may cause contamination and degradatio…
- Class IIOngoing2022-06-13There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to leave the blastocyte to air thaw for 5 minutes rather than the correct 5 seconds as indicated in the English version.
- Class IIOngoing2022-04-01Products in this lot may have compromised barrier seals, introducing a risk to sterility.
- Class IIOngoing2022-03-31Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure
20215 records
- Class IIOngoing2021-09-29XXX
- Class IITerminated2021-06-30Units are displaying output readings that differ from the measurements taken using a Gas Analyzer. When the display is reading a target setting of 5%, an external meter is reading between 3% and 12% and the Incubator does not alarm. May result in loss of embryo during storage.
- Class IITerminated2021-04-30The affected Wallach LOOP Electrodes were incorrectly packaged with a Round Electrode inside a package labeled as a Square Electrode.
- Class IIOngoing2021-04-23The affected Koh Cup were incorrectly packaged with a 3.5cm Ultem Koh cup, P/N 34131-04, instead of a 3.0cm Ultem Koh cup, P/N 34131-03 as required.
- Class IITerminated2021-01-21The affected Milex Gellhorn pessaries were incorrectly manufactured with a long stem (2.170 ), instead of a short stem (1.670 or 1.960 nominal height) as required. The length discrepancy does not meet the released print dimensions.
20204 records
- Class IITerminated2020-09-08There is a discrepancy between the printing on the Product Tyvek label of the individual component packaging. The Chinese characters on the label indicates that it contains an Equilibration Medium (P/N: 12220001) rather than the Vitrification Medium (P/N: 12230001) as indicated o…
- Class IITerminated2020-06-03Mislabeled: Printing on the top web or Tyvek lid of the individual product blister tray indicating that it contains a Pipet Current or Part Number MX140 and the display box indicating the correct Os FinderTM Cervical Dilators, Part Number 1176. The product in the individual prod…
- Class IITerminated2020-04-09The seal of the sterile pouch may be compromised on one lot of the Milex Cervical Dilators.
- Class IIITerminated2020-02-28Expiration date discrepancy. Marked as 5 years. Expiration date should be 3 years.
20199 records
- Class IITerminated2019-12-13Potential for unsealed pouches in two (2) lots of the Endosee Advance Cannula (P/N: ESPX5).
- Class IITerminated2019-10-16Potential for the Quantum 2000 Electrosurgery Generator, part numbers 909075, 909075-05 and 909075-05W, to cease to either cut, coagulate, or blend during a procedure when the foot pedal is used for activation.
- Class IIITerminated2019-09-05The touchpad may not work properly after cleaning.
- Class IIITerminated2019-08-19Discrepancy between the correct expiration dates on the individual pouches inside the boxes and incorrect expiration dates on the outer boxes, where the month and day were reversed.
- Class IITerminated2019-07-08The firm updated the IFU in April 2019 to clarify that use of the TRANSWARMER Mattress with other heat producing devices, such as an incubator, is prohibited and could lead to serious health consequences, such as skin burns. some products do not contain the updated IFU.
- Class IITerminated2019-06-18The products' cut, coagulate, or blend function may not operate with the use of the Foot Pedal.
- Class IITerminated2019-06-14The soft, flexible cup of the Advincula Delineator may be unvalidated and could potentiall crack when subjected to excessive compressions during use, leading to serious health consequences, such as lacerations.
- Class IITerminated2019-04-232.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses.
- Class IITerminated2019-03-27ICSI CUSTOM MICROPIPETTES the thirty-degree bend is absent
20184 records
- Class IITerminated2018-09-04There is a possibility that the seal of the sterile pouch may be compromised, thereby increasing the risk of infection.
- Class IITerminated2018-09-04The seal of the sterile pouch may be compromised, thereby increasing the risk of infection.
- Class IITerminated2018-03-12Products have been packaged with an incorrect product size.
- Class IITerminated2018-03-12Products have been packaged with the wrong Trocar length type.
20172 records
- Class IITerminated2017-10-06The 2MHZ prove was incorrectlhttp://cts.fda.gov/division-tracking/images/trash.pngy labeled as a 3MHZ probe
- Class IITerminated2017-04-10The current secondary packaging, a box, incorrectly states that the product is latex free. The product is made with 100% latex. The label on the package that contains the product and the Instructions for Use (IFU) correctly states that the pessary is manufactured from latex rubbe…
20162 records
- Class IIITerminated2016-08-25Products contain international package insert and not the correct USA package insert
- Class IITerminated2016-02-16The products have been identified to contain an incorrect size Diaphragm.
201511 records
- Class IIITerminated2015-09-14The products have been identified to have incorrect size prints on the MILEX PESSARY KIT packaging. While the Pessary pouch is marked with correct size, an incorrect label on the packaging does not meet the required release specification.
- Class IIITerminated2015-09-01Kit Label incorrect :Labeled with Lidocaine HCI drug cartridges Lidocaine HCl (2%) but contains vials Lidocaine HCl (1%)
- Class IITerminated2015-07-06Product marketed without a 510(k)
- Class IITerminated2015-06-19Incorrect expiration date on outer carton kit label
- Class IITerminated2015-05-20A diaphragm size 80 labeled box was incorrectly packaged with a size 85.
- Class IITerminated2015-05-18Design of device may expose user to injury to fingers or body parts
- Class IITerminated2015-04-13Small sized specula were incorrectly packaged in printed bags for the medium sized speculum
- Class IITerminated2015-03-20The pad printing around the suture entry holes on the 15mm Suture Guide included in the kit may not have been appropriately manufactured resulting in the pad printing flaking off around the entry holes.
- Class IITerminated2015-03-09Internal gas line may malfunction causing the line to burst under pressure.
- Class IITerminated2015-02-09Incorrect size printed on the Milex Pessary Kit packaging label
- Class IITerminated2015-01-26Products without expiration date have the remote potential for minor burn
20145 records
- Class IIITerminated2014-10-27The product has been identified to contain an incorrect curette type.
- Class IITerminated2014-10-27Outer box is mislabeled as P/N R2010- Large Radius Loop Electrode instead of P/N R1010- Medium Radius Loop Electrode
- Class IIITerminated2014-09-30Incorrect expiration date printed on the MILEX PESSARY KIT packaging
- Class IITerminated2014-07-03Unsealed pouches.
- Class IITerminated2014-05-13Sterility of the device may be compromised due to unsealed pouch
20131 record
- Class IITerminated2013-05-01Trans Warmer Infant and Transport Mattress not labeled with exp.date may not meet specification
Related
- All facilities operated by COOPERSURGICAL, INC.
- Lone Star Medical Products (TX)
- CooperSurgical Distribution B.V. (NL)
- COOPERSURGICAL, INC. (CT)
- HBH MEDIZINTECHNIK GMBH (DE)
- LEISEGANG FEINMECHANIK-OPTIK GMBH (DE)
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “COOPER MEDICAL, SRL (FEI 3014295858) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3014295858-cooper-medical-srl/
Alerts for COOPER MEDICAL, SRL
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.