RegWardenFDA facility compliance records
Recall

Class II — Baxter Healthcare Corporation

Recalling firmBAXTER HEALTHCARE CORPORATION
Registered facilityView this company's 32 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonPotential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue.
ProductLikorall Overhead Lifts Product Code: 3121001 Product Description: LIKORALL 200 Product Code: 3122005 Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122005NA Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122006 Product Description: LIKORALL 242 ES, WHITE Product Code: 3122006NA Product Description: LIKORALL 242 ES, WHITE Product Code: 3122007 Product Description: LIKORALL 242 ES R2R, NATURAL Product Code: 3122007NA Product Description: LIKORALL 242 ES R2R, NATURAL Product Code: 3122008 Product Description: LIKORALL 242 ES R2R, W…
DistributionWorldwide - US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina…
Initiated2026-09-14
Classified2026-10-13
Recall numberZ-0151-2027
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-29. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Baxter Healthcare Corporation (Z-0151-2027).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-29. https://regwarden.com/recall/99802/

Not open yet

Alerts for BAXTER HEALTHCARE CORPORATION

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.