RegWardenFDA facility compliance records
Recall

Class II — IMRIS Imaging Inc

Recalling firmIMRIS Imaging, Inc.
Registered facilityIMRIS Imaging, Inc. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThe linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
ProductORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000
Quantity6 units
DistributionWorldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.
Initiated2026-08-06
Classified2026-08-26
Recall numberZ-3033-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — IMRIS Imaging Inc (Z-3033-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/99596/

Not open yet

Alerts for IMRIS Imaging, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.