Class II — Medline Industries, LP
| Recalling firm | Medline Industries, LP |
|---|---|
| Registered facility | View this company's 31 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening. |
| Product | Medline Convenience Kits: 1) BLOOD DRAW KIT, Catalog Number: DYLAB1081; 2) WHOLE BLOOD PHLEBOTOMY KIT, Catalog Number: DYNDB3005D; 3) QP KIT, Catalog Number: DYNDB3005F; 4) DOUBLE RED CELL PHLEBOTOMY KIT, Catalog Number: DYNDB3006C; 5) PLATELET PHLEBOTOMY KIT, Catalog Number: DYNDB3007C; 6) WHOLE BLOOD KIT, Catalog Number: DYNDB3041D; 7) PHLEBOTOMY, Catalog Number: DYNDB3044D; 8) PHLEBOTOMY, Catalog Number: DYNDB3044F; 9) DONATION - WB, Catalog Number: DYNDB3086D; 10) DONATION - ALYX/TRIMA W/O TOUR, Catalog Number: DYNDB3088B; 11) THERAPEUTIC NFT, Catalog Number: DYNDB3100… |
| Quantity | 996692 kits |
| Distribution | US Nationwide distribution. |
| Initiated | 2026-06-12 |
| Classified | 2026-08-31 |
| Recall number | Z-3063-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Medline Industries, LP (Z-3063-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/99334/
Alerts for Medline Industries, LP
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.