RegWardenFDA facility compliance records
Recall

Class II — Permobil Inc

Recalling firmPermobil (KunShan) Co., Ltd
Registered facilityPERMOBIL (KUNSHAN) CO., LTD (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThis recall has been reported retroactively. The power wheelchair firmware lacks a speed restriction to maintain stability when the seat lift exceeds 50 mm, which may cause the user to fall from the wheelchair or the wheelchair to tip over,
ProductPermobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model version 3
Quantity140 units
DistributionWorldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CAN.
Initiated2024-02-29
Classified2026-07-24
Recall numberZ-2795-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Permobil Inc (Z-2795-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/99315/

Not open yet

Alerts for Permobil (KunShan) Co., Ltd

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.