Class II — Boston Scientific Corporation

Recalling firmBoston Scientific Corporation
Registered facilityView this company's 37 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonBoston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
ProductFARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01
Quantity508,151 units
DistributionGlobal Distribution. US Nationwide.
Initiated2026-06-11
Classified2026-07-17
Recall numberZ-2756-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Boston Scientific Corporation (Z-2756-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/99252/

Not open yet

Alerts for Boston Scientific Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.