RegWardenFDA facility compliance records
Recall

Class II — Think Surgical, Inc.

Recalling firmThink Surgical Inc.
Registered facilityThink Surgical Inc. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonDue to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array.
ProductTMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI code: M9461090040 TMINI Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components
Quantity360 devices
DistributionU.S. Nationwide distribution in the states of AZ, CA, DE, FL, IA, KS, LA, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OR, PA, TX, UT, VA, and WA.
Initiated2026-05-08
Classified2026-07-09
Recall numberZ-2716-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Think Surgical, Inc. (Z-2716-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/99219/

Not open yet

Alerts for Think Surgical Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.