RegWardenFDA facility compliance records
Recall

Class II — PFM MEDICAL INC.

Recalling firmPFM Medical
Registered facilitypfm medical cpp SA (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonA snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.
ProductMulti-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310
Quantity32
DistributionUS Nationwide distribution in the states of NY, MD.
Initiated2026-04-21
Classified2026-06-23
Recall numberZ-2526-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — PFM MEDICAL INC. (Z-2526-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98942/