Recall

Class II — Mint Medical GmbH

Recalling firmMint Medical GmbH
Registered facilityMINT MEDICAL GMBH (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonIf the connection between a mint Lesion workstation and the mint Lesion server is interrupted while performing a read, in rare cases some information may be lost or incorrectly linked to other patients.
ProductMint Lesion; Software Versions: 3.4.0 up to 3.9.5.;
Quantity101 systems
DistributionWorldwide distribution - US Nationwide and the countries of Germany, Switzerland, Austria, Belgium, Italy, France, Romania, Netherlands, United Kingdom, Spain, Czech Republic, Ireland, Poland;
Initiated2026-01-07
Classified2026-05-06
Recall numberZ-2068-2026
InitiationVoluntary: Firm initiated