RegWardenFDA facility compliance records
Recall

Class II — Smith & Nephew, Inc.

Recalling firmSmith & Nephew, Inc.
Registered facilityView this company's 15 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonHumeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.
ProductTRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
Quantity5
DistributionInternational distribution to the countries of BE, DE, IT.
Initiated2026-03-16
Classified2026-06-16
Recall numberZ-2456-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Smith & Nephew, Inc. (Z-2456-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98629/