RegWardenFDA facility compliance records
Recall

Class II — ICU Medical, Inc.

Recalling firmICU MEDICAL, INC.
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonOncology sets/closed system transfer devices with extension sets may leak resulting in therapy delayed if noticed and corrected during setup, but if the leak occurs during therapy this may result in therapy interruption with the potential for the patient to receive less infusion solution than intended. Exposure of hazardous substances to the patient, caregiver, or pharmacy technician is possible.
ProductOncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960
Quantity20,506
DistributionUS Nationwide distribution in the states of GA, MA, OH, NC, CA, ND, MI, TX, FL, AR, CT, PA, WA, WI, CO, VT, NY, NH, KY, IA, KS, AL, NM, IL, MO, NJ.
Initiated2026-03-24
Classified2026-05-18
Recall numberZ-2210-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — ICU Medical, Inc. (Z-2210-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98622/

Not open yet

Alerts for ICU MEDICAL, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.