RegWardenFDA facility compliance records
Recall

Class II — Beckman Coulter, Inc.

Recalling firmBECKMAN COULTER, INC.
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonBeckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.
ProductAccess Total T4 Calibrator, Catalog No. 33805
Quantity2068 US, 2830 OUS
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Austria, Bangladesh, Bolivia (Plurinational State of), Bosnia and Herzegovina, Brazil, Canada, China, Cte d'Ivoire, Croatia, Czechia, Denmark, Egypt, El Salvador, Ethiopia, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Kosovo, Lebanon, Libya, Malawi, Mexico, Myanm…
Initiated2026-03-16
Classified2026-04-22
Recall numberZ-1922-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Beckman Coulter, Inc. (Z-1922-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98591/

Not open yet

Alerts for BECKMAN COULTER, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.