Class II — Hologic, Inc
| Recalling firm | HOLOGIC, INC. |
|---|---|
| Registered facility | View this company's 12 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Complaints have been received of systems developing loose, missing, or broken internal bolts over time. |
| Product | 3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-INTL3D, 3DM-SYS-INTL3D-MOB, 3DM-SYS-INTL3D-NS, 3DM-SYS-STD, 3DM-SYS-STD-MOB, 3DM-SYS-STD-NS. |
| Quantity | 14837 units |
| Distribution | US Nationwide distribution. |
| Initiated | 2026-04-08 |
| Classified | 2026-05-12 |
| Recall number | Z-2168-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)