RegWardenFDA facility compliance records
Recall

Class II — Spacelabs Healthcare, Inc.

Recalling firmSPACELABS HEALTHCARE, INC.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonMulti-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
ProductMulti-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Quantity1790
DistributionWorldwide - US Nationwide distribution including in the states of NY, NJ, AL, OK, IA, MT, WY, MN, TX, WI, PA, WA, OR, MI, CA, SC, ID, FL, MS, KS, SD, LA, OH, MO, NH, GA, MD, TN, AZ, PR and the countries of TAIWAN, CANADA, KUWAIT, FRANCE, SAUDI ARABIA, POLAND, PAKISTAN, MEXICO, ARGENTINA, PANAMA.
Initiated2026-02-27
Classified2026-04-07
Recall numberZ-1793-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Spacelabs Healthcare, Inc. (Z-1793-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98374/

Not open yet

Alerts for SPACELABS HEALTHCARE, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.