RegWardenFDA facility compliance records
Recall

Class II — Ion Beam Applications S.A.

Recalling firmION BEAM APPLICATIONS S.A.
Registered facilityION BEAM APPLICATIONS S.A. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonIt was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
ProductIBA Proton Therapy System - PROTEUS 235
Quantity4 units (1 US, 3 OUS)
DistributionWorldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.
Initiated2026-02-03
Classified2026-03-06
Recall numberZ-1497-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Ion Beam Applications S.A. (Z-1497-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98364/

Not open yet

Alerts for ION BEAM APPLICATIONS S.A.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.