RegWardenFDA facility compliance records
Recall

Class II — ETAC A/S

Recalling firmEtac A/S
Registered facilityEtac A/S (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonA customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
ProductBrand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M15001, (7) M15002; Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength. Component: The deformation occurs between the bolt and the f…
Quantity71
DistributionWorldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.
Initiated2025-11-17
Classified2026-02-05
Recall numberZ-1309-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — ETAC A/S (Z-1309-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/98087/

Not open yet

Alerts for Etac A/S

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.