Class II — Remote Diagnostic Technologies Ltd.
| Recalling firm | REMOTE DIAGNOSTIC TECHNOLOGIES LTD. |
|---|---|
| Registered facility | REMOTE DIAGNOSTIC TECHNOLOGIES LTD. (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Patient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for accuracy. If healthcare providers base clinical judgment on unvalidated measurements, this could potentially lead to incorrect clinical decisions. |
| Product | Philips Tempus Pro Patient Monitor, REF: 00-1004, 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R. USB 2-Channel IBP Module, REF: 01-2017. Used with User/Operator Manual |
| Quantity | 7129 |
| Distribution | Worldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE, IL, SG, CA, IE, TH, NL, MY, AT, DK, RO, CH, IS, PT, NO, HK, PL, KY, OM, CZ, SI, SL, GI, ZA, AF, SE, FI. |
| Initiated | 2025-11-26 |
| Classified | 2026-01-08 |
| Recall number | Z-1027-2026 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Remote Diagnostic Technologies Ltd. (Z-1027-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97889/
Alerts for REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.