Recall

Class II — Inpeco S.A.

Recalling firmInpeco SA
Registered facilityINPECO SA (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThe Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes.
ProductFlexLab (FLX); Version: FLX-217-10;
Quantity1 unit (OUS only)
DistributionUS Domestic: MI. OUS International: Australia, Finland, France, Germany, Italy, Portugal, Spain, and UK.
Initiated2025-10-30
Classified2025-11-25
Recall numberZ-0595-2026
InitiationVoluntary: Firm initiated