Class I — Siemens Medical Solutions USA, Inc

Recalling firmSIEMENS MEDICAL SOLUTIONS USA, INC.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusOngoing
ReasonThere is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
ProductMAGNETOM Cima.X Upgrade. Model Number: 11689304.
Quantity4 system (1 US; 3 OUS)
DistributionUS distribution to California. International distribution to Australia, Germany, Great Britian.
Initiated2025-10-01
Classified2025-11-12
Recall numberZ-0431-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Siemens Medical Solutions USA, Inc (Z-0431-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97845/

Not open yet

Alerts for SIEMENS MEDICAL SOLUTIONS USA, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.