RegWardenFDA facility compliance records
Recall

Class II — Spacelabs Healthcare, Inc.

Recalling firmSPACELABS HEALTHCARE, INC.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonDue to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.
ProductBrand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-01 Software Version: N/A Product Description: 6 foot DVI-I to DVI-I male/male display cable. Component: Yes, the cable is an external display cable used with the 91390 and 91393
Quantity8179
DistributionWorldwide - United States Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of AFG, ARE, ARG, AUS, AUT, BGD, BHR, BOL, BRA, BRN, CAN, CHE, CHL, CHN, COL, CRI, CYP, CZE, DOM, DZA, ECU, EGY, ESP, FRA, GBR, GR…
Initiated2025-09-05
Classified2025-10-16
Recall numberZ-0167-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Spacelabs Healthcare, Inc. (Z-0167-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97616/

Not open yet

Alerts for SPACELABS HEALTHCARE, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.