Class I — Siemens Medical Solutions USA, Inc
| Recalling firm | SIEMENS MEDICAL SOLUTIONS USA, INC. |
|---|---|
| Registered facility | View this company's 9 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class I |
| Status | Ongoing |
| Reason | There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room. |
| Product | MAGNETOM Skyra fit. Model Number: 10849580. |
| Quantity | 56 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyprus, Czechia, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Ethiopia, Finland, Faroe Is… |
| Initiated | 2025-08-28 |
| Classified | 2025-09-30 |
| Recall number | Z-2661-2025 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class I recall — Siemens Medical Solutions USA, Inc (Z-2661-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97530/
Alerts for SIEMENS MEDICAL SOLUTIONS USA, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.