RegWardenFDA facility compliance records
Recall

Class II — Zap Surgical Systems

Recalling firmZap Surgical Systems Inc.
Registered facilityZAP SURGICAL SYSTEMS INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonIf the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.
ProductZAP-X Radiosurgery System, Model: 300150
Quantity17
DistributionWorldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, Trkiye, Paraguay.
Initiated2025-07-25
Classified2025-09-26
Recall numberZ-2639-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Zap Surgical Systems (Z-2639-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97413/

Not open yet

Alerts for Zap Surgical Systems Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.