Recall

Class II — PIE Medical Imaging B.V.

Recalling firmPie Medical Imaging B.V.
Registered facilityPIE MEDICAL IMAGING B.V. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonWhen fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock measurements are not updated relative to the new 12h position which may cause stent graft fenestrations at incorrect position, which may lead to blood flow disruption and tissue damage
Product3mensio Workstation (Vascular Fenestrated) software
Quantity501
DistributionUS Nationwide distribution in the states of MN, NJ, GA, NY, OH, KS, MA, CA, TX, MS, NE, NC.
Initiated2025-07-14
Classified2025-08-12
Recall numberZ-2297-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — PIE Medical Imaging B.V. (Z-2297-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97307/

Not open yet

Alerts for Pie Medical Imaging B.V.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.