RegWardenFDA facility compliance records
Recall

Class II — Siemens Healthcare Diagnostics Inc

Recalling firmSIEMENS HEALTHCARE DIAGNOSTICS INC.
Registered facilityView this company's 12 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonSiemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.
ProductRAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
Quantity1,477 units
Distributionworldwide distribution - US Nationwide and the countries of Canada, Chile, Japan, Paraguay.
Initiated2025-06-26
Classified2025-07-31
Recall numberZ-2230-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Healthcare Diagnostics Inc (Z-2230-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/97209/

Not open yet

Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.