RegWardenFDA facility compliance records
Recall

Class I — Maquet Critical Care AB

Recalling firmMaquet Critical Care AB
Registered facilityMAQUET CRITICAL CARE AB (the firm's only registered facility)
ClassificationClass I
StatusOngoing
ReasonPotential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.
ProductServo-u MR Ventilator System. Model Number: 6888800.
Quantity129 units
DistributionUS Nationwide distribution and Puerto Rico. No international distribution.
Initiated2025-05-30
Classified2025-06-27
Recall numberZ-2015-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)