Class II — ICU Medical, Inc.
| Recalling firm | ICU MEDICAL, INC. |
|---|---|
| Registered facility | View this company's 6 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection. |
| Product | SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000 |
| Quantity | 1,445,600 |
| Distribution | Worldwide - US Nationwide distribution including in the states of TX, SD, ID, LA, IN, CO, OK, MO, MN, OR, AZ, NC, TN, VA, CA, IL, FL, MA, OH, MD, PA, NJ, NY, WA, GA, UT, MI, MS, AR, AL, NV, KS, IA, WV, HI, KY, NM and the country of CA. |
| Initiated | 2025-06-03 |
| Classified | 2025-07-11 |
| Recall number | Z-2116-2025 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — ICU Medical, Inc. (Z-2116-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96903/
Alerts for ICU MEDICAL, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?