RegWardenFDA facility compliance records
Recall

Class II — VIRTUAL INCISION CORPORATION

Recalling firmVirtual Incision Corporation
Registered facilityVIRTUAL INCISION CORPORATION (the firm's only registered facility)
ClassificationClass II
StatusCompleted
ReasonThe MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.
ProductMIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3
Quantity3 units
DistributionUS Distribution to States: CA, FL, NE
Initiated2025-04-23
Classified2025-06-04
Recall numberZ-1905-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — VIRTUAL INCISION CORPORATION (Z-1905-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96831/

Not open yet

Alerts for Virtual Incision Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.