Class II — IHB OPERATIONS B.V.

Recalling firmIHB operations
Registered facilityIHB OPERATIONS (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThe mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
ProductSkytron Freedom 4FXS-60 Heavy Duty Monitor Mount
Quantity112 units
DistributionUS Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
Initiated2025-03-27
Classified2025-04-23
Recall numberZ-1663-2025
InitiationVoluntary: Firm initiated