Class II — IHB OPERATIONS B.V.
| Recalling firm | IHB operations |
|---|---|
| Registered facility | IHB OPERATIONS (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column. |
| Product | Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount |
| Quantity | 112 units |
| Distribution | US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV. |
| Initiated | 2025-03-27 |
| Classified | 2025-04-23 |
| Recall number | Z-1663-2025 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — IHB OPERATIONS B.V. (Z-1663-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96581/
Alerts for IHB operations
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.