Class II — Beckman Coulter, Inc.

Recalling firmBECKMAN COULTER, INC.
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonWhen DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.
ProductBeckman Coulter DxI 9000 Access Immunoassay Analyzer
Quantity370 units
DistributionWorldwide distribution.
Initiated2025-03-27
Classified2025-04-24
Recall numberZ-1664-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Beckman Coulter, Inc. (Z-1664-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96554/

Not open yet

Alerts for BECKMAN COULTER, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.