RegWardenFDA facility compliance records
Recall

Class II — MEDLINE INDUSTRIES, LP - Northfield

Recalling firmMedline Industries, LP
Registered facilityView this company's 31 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonPotential breach in pouch packaging which could lead to loss of sterility.
ProductProxima Drape labeled as UNIVERSAL DAVINCI ROBOTIC P. Sterile surgical drape.
Quantity220 units
DistributionUS Nationwide distribution.
Initiated2025-02-13
Classified2025-03-14
Recall numberZ-1369-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — MEDLINE INDUSTRIES, LP - Northfield (Z-1369-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96401/

Not open yet

Alerts for Medline Industries, LP

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.