Recall

Class II — TAS Medical Inc

Recalling firmTAS Medical, Inc.
Registered facilityTAS MEDICAL, INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonTissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.
ProductTISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP
Quantity114
DistributionUS Nationwide distribution in the states of LA, OK, TN.
Initiated2025-01-27
Classified2025-03-06
Recall numberZ-1294-2025
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — TAS Medical Inc (Z-1294-2025).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/96274/

Not open yet

Alerts for TAS Medical, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.