Class II — Beckman Coulter, Inc.
| Recalling firm | BECKMAN COULTER, INC. |
|---|---|
| Registered facility | View this company's 22 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results. |
| Product | DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137. |
| Quantity | 206 units |
| Distribution | US Nationwide. Global Distribution. |
| Initiated | 2025-02-05 |
| Classified | 2025-03-17 |
| Recall number | Z-1388-2025 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)